MJQ-30
Point of origin
China CE
Throughput
Detection mode: 21 minutes; standard mode: 35 minutes
Application scenarios
Hospital Endoscopy Center Operating Room

MJQ-30 Hydrogen peroxide low-temperature plasma sterilizer

  1. ★The total design volume of the sterilization chamber is ≥42.9L, and the solvent used is 28-30L.

  2. Sterilization chamber: rectangular (square); the inner chamber is made of aviation aluminum, with a thickness of ≥8mm.

  3. Sealed door: single door, made of aviation aluminum, with a thickness of ≥20mm.

  4. Filling method: cassette type; each cassette has ≥12 capsules, and the hydrogen peroxide capacity of a single capsule is ≤2ml.

  5. Capsule counting memory function: After the cassette is installed, the number of capsules used is automatically calculated, and the number of remaining capsules and the number of full cycles that can be run are prompted.

  6. Automatic identification system: Automatically identify whether the cassette model matches the equipment, and automatically identify the validity period of the cassette.

  7. ★The concentration of hydrogen peroxide sterilant is ≤59.5%, and has the qualification for dangerous chemical business license

  8. Amount of sterilant: detection mode: ≤2mL/time, standard mode: ≤4mL/time.

  9. ★The evaporator is equipped with a concentration purification device, and the concentration of hydrogen peroxide after purification is ≥97%

  10. Sterilization program and sterilization time: detection mode ≤23 minutes, standard mode ≤35 minutes, and with program countdown function.

  11. ★Control system: PLC

  12. Display: ≥7-inch TFT true color touch screen.

  13. Recording method: micro printer prints records.

  14. Data storage function: can store sterilization process data for at least 3 years or 30,000 cycles, which can be retrieved and printed at any time.

  15. Can automatically monitor plasma operating power, pressure data of each stage, time of each stage, and temperature of the entire operation.

  16. Protection device: equipped with power protector, with the function of automatically monitoring whether the user's power supply is overvoltage, undervoltage, or phase sequence error; equipped with automatic fault detection, with fault code display alarm, fault audio alarm and fault recording function; the sterilizer sealed door and sterilization chamber are equipped with temperature protection devices respectively, which can automatically perform preheating control; equipped with automatic obstacle avoidance protection function for sealing door closing.

  17. Information interface: can be connected to the CSSD traceability system.

  18. Pipeline: SUS304 stainless steel or higher specification pipe.

  19. Vacuum pump: adopts a rotary vane vacuum pump resistant to hydrogen peroxide corrosion and equipped with a vacuum pump phase sequence protector.

  20. Convenient operation: the cavity storage basket can be adjusted to 1~2 layers or more.

  21. Biological monitoring bacteria: thermophilic Bacillus spores and Bacillus subtilis black spores.

  22. ★ It can sterilize single-channel stainless steel catheters with an inner diameter of ≥0.7mm and a length of ≤600mm or single-channel polytetrafluoroethylene catheters with an inner diameter of ≥1mm and a length of ≤4000mm.

  23. ★ The residual hydrogen peroxide in the polytetrafluoroethylene lumen after sterilization is less than 0.0008mg/cm2, and the residual hydrogen peroxide in the stainless steel lumen is less than 0.0008mg/cm2.

  24. The evaluation of the compatibility of metal and non-metal instruments should lead to the conclusion that there is no corrosion and no impact on the expected clinical service life.

  25. ★ The residual value of hydrogen peroxide in the air for 8 hours is ≤0.6mg/m3.

  26. ★ There is a test report on the accuracy of the "Tianzhihang Orthopedic Surgery Positioning and Navigation Tool Kit" that can still achieve normal use after sterilization 125 times.

  27. CE certification


The MJQ-30 Hydrogen Peroxide Low-temperature Plasma Sterilizer is designed for a range of clinical applications, particularly for the sterilization of medical instruments and devices that are sensitive to heat and moisture. Its advanced features make it suitable for various clinical settings, ensuring effective sterilization while preserving the integrity of the instruments.

Clinical Applications

  1. Surgical Instruments:

    • Ideal for sterilizing complex and delicate surgical instruments, including those made of metals and non-metals.

    • Ensures no corrosion and maintains the expected clinical service life of the instruments.

  2. Endoscopic Equipment:

    • Suitable for sterilizing endoscopes and related accessories, which are often heat-sensitive and cannot be sterilized using traditional high-temperature methods.

  3. Catheters:

    • Capable of sterilizing single-channel stainless steel catheters with an inner diameter of ≥0.7mm and a length of ≤600mm.

    • Also suitable for single-channel polytetrafluoroethylene (PTFE) catheters with an inner diameter of ≥1mm and a length of ≤4000mm.

  4. Orthopedic Surgery Tools:

    • Proven to maintain the functionality of orthopedic surgery positioning and navigation tool kits even after repeated sterilization (125 cycles as per test reports).

  5. Dental Instruments:

    • Effective for sterilizing dental handpieces, mirrors, and other tools, ensuring they are free from microbial contamination.

  6. Reusable Medical Devices:

    • Suitable for various reusable medical devices, ensuring they are thoroughly sterilized without damage.

Hydrogen peroxide low temperature plasma sterilizer Performance parameters list
modelMJQ-30
Effective volume28L-30L
Power supply3N~380V 50Hz
Working systemShort-time loading, continuous operation
Sterilization time (excluding vacuum detection time)Detection mode: 21 minutes; standard mode: 35 minutes
Input power2700VA
Constant temperature power700W
Standby powerAbout 25W
Operating environment temperature5C~40
Operating environment humidity30%~95% (no condensation)
Operating environment atmospheric pressure700hPa~1060hPa
Installation methodDesktop installation (optional bracket)
58% hydrogen peroxide dosage (mL)Detection mode: 2mL/time Standard mode: 4mL/time
Number of uses per cartridgeDetection mode: 12 times Standard mode: 6 times
Sterilization chamber operating temperature45'C~60C
Minimum vacuum degree during sterilization chamber operation≤80Pa
Sterilization chamber shapeRectangular
Overall dimensions (width x depth x height)720X 990X 810mm 
Packaging dimensions (width x depth x height )833 X 1087X1037mm 
Minimum dimensions required for equipment installation (width x depth x height )1120X 1500X 2000mm 
Sterilization chamber dimensions (width x depth x height)260X 750X220mm 
Storage bar dimensions (width x depth)200X710mm 
Number of storage bars2 layers
Storage bar load capacity5Kg/layer
Net weight of the whole machine215Kg
Gross weight of the whole machine275Kg
Sterilization chamber materialAluminum alloy
Casing materialQ235
Theoretical maintenance cycle of the equipmentHalf a year
Display7-inch TFT true color ribbon touch screen function
Data recording and queryAny time query and print
PrinterDot matrix printer


×